The analysis of cargo proteins in exosome subpopulations has considerable value in diagnostics but a translatable impact has been limited by lengthy or complex exosome extraction protocols. We describe herein a scalable, fast, and low-cost exosome extraction using an alternating (AC) magnetic field to support the dynamic mixing of antibody-coated magnetic beads (MBs) with serum samples within 3D-printed microfluidic chips. Zwitterionic polymer-coated MBs are, specifically, magnetically agitated and support ultraclean exosome capture efficiencies >70% from <50 μL of neat serum in 30 min. Applied herein to the immunocapture of neuronal exosomes using anti-L1CAM antibodies, prior to the array-based assaying of α-synuclein (α-syn) content by a standard duplex electrochemical sandwich ELISA, sub pg/mL detection was possible with an excellent coefficient of variation and a sample-to-answer time of ∼75 min. The high performance and semiautomation of this approach hold promise in underpinning low-cost Parkinson's disease diagnostics and is of value in exosomal biomarker analyses more generally.

More information

DOI

10.1021/acs.analchem.3c00357

Type

Journal article

Publication Date

23/05/2023

Volume

95

Pages

7906 - 7913

Total pages

7

Keywords

Humans, Parkinson Disease, Exosomes, Magnetic Fields, Microfluidics